Z-2192-2025Class II
Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35
Phasor Health, LLC / initiated 2025-03-21 / Open, Classified
Phasor Health, LLC began a recall of Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35 on . FDA classified it as Class II and gives the reason as "Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2192-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Phasor Health, LLC
- Product
- Phasor Drill, REF: 2.70S, 3.20L, 4.50, 5.30, 6.35
- Product code
- HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Reason for recall
- Drill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time.
- FDA root cause
- Process change control
- Action
- Starting on 3/21/2025, device correction and removal began and a correction and removal notice was emailed. Customers were asked to do the following: 1) Review inventory (without opening the sterile transparent barrier) to determine if any of the affected lots remain in inventory. If yes, remove those lots from inventory. Do not use drills with reverse battery orientation. 2) If unable to review inventory soon, the Company will come to your site within 30 business days of this letter and inspect your inventory (within 20 minutes total) and replace any defective drills with new verified stock (same catalog number). 3) Contact firm at [email protected] to report results of inventory inspection and to facilitate device return. 4) Complete and return the acknowledgement and receipt form via email to [email protected] Please contact Firm at [email protected] or (832) 982-1234 for more information or questions
- Quantity in commerce
- 1,064
- Distribution
- US: PA, IN, AZ, MO, LA, FL, TX, MA, OH, TN, MI, WV, VA, CA, DE, NH, GA, SC, GU, NY, NJ, MN, WI, NE, WA, IL, AR, PR, VT
- Product codes and lots
- UDI-DI: B775270S0, B775320L0, B7754500, B7755300, B7756350. Lot: 241002450, 241002530, 241009450, 241009530, 241016450, 241016530, 241023450, 241023530, 241030450, 241030530, 24112027S, 24121827S, 24122727S, 25012227S, 24082832L, 24102332L, 241227450, 241220635, 250122635, 241227530, 250122530.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2192-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.