Z-2292-2025Class II

Bartels ELISA Legionella Urinary Antigen. Model Number: B1029-440. In Vitro Diagnostic test kit for the...

Trinity Biotech USA / initiated 2025-07-22 / Open, Classified

Trinity Biotech USA began a recall of Bartels ELISA Legionella Urinary Antigen. Model Number: B1029-440. In Vitro Diagnostic test kit for the presumptive diagnosis of past or current Legionnaires... on . FDA classified it as Class II and gives the reason as "Potential for Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 065 and 066 to contain particulate matter". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2292-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Trinity Biotech USA
Product
Bartels ELISA Legionella Urinary Antigen. Model Number: B1029-440. In Vitro Diagnostic test kit for the presumptive diagnosis of past or current Legionnaires Disease.
Product code
MJH Legionella, Spp., Elisa
Reason for recall
Potential for Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 065 and 066 to contain particulate matter.
FDA root cause
Under Investigation by firm
Action
Trinity Biotech initially notified consignees 06/05/2025 via letter. Consignees were instructed to inspect the conjugate bottles before use in both manual and automated testing and if any are observed with particulate matter, Trinity Biotech confirms the kit performs, and results are consistent with release specifications and IFU claims. However, replacement can be requested. Trinity Biotech notified consignees again on about 07/22/2025 via an updated URGENT: MEDICAL DEVICE Notification letter. Consignees were instructed that their trained laboratory technicians should inspect the conjugate bottles before use in both manual and automated testing, if any bottle is observed to contain conjugate with particulate matter, to discard the bottle and use a clear bottle available. If there are no other available clear bottles, please contact Trinity Biotech for a replacement kit. The current conjugate will be replaced with Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 067 once available.
Quantity in commerce
968 kits
Distribution
Worldwide - US Nationwide distribution in the states of Maryland and Utah. The countries of France, Germany, Ireland, Israel, Kuwait, Latvia, UK.
Product codes and lots
Model Number: B1029-440. UDI-DI: 05391516744065. Lot Numbers: 065 and 066

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2292-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2292-2025Class IIOngoingVoluntary: Firm initiated