Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1...
Medtronic, Inc. began a recall of Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number... on . FDA classified it as Class II and gives the reason as "In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2168-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;
- Product code
- NVZ Pulse Generator, Permanent, Implantable
- Reason for recall
- In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- An URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letter dated 6/18/25 was sent to customers. Customer Actions: "Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronicrepresentative for update instructions. If needed, Medtronic representatives are available to work with you toinstall or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify thesoftware update is complete. "Sign and return the enclosed confirmation form and share this notice with those who need to be aware withinyour organization or with any organization where SmartSync may be in use. Questions regarding this informationshould be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.
- Quantity in commerce
- 22091 total devices
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- 1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2168-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2169-2025 | Class II | Ongoing | Voluntary: Firm initiated |