Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP...
Stryker Corporation began a recall of Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP... on . FDA classified it as Class II and gives the reason as "Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2277-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235
- Product code
- KCY Tourniquet, Pneumatic
- Reason for recall
- Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- On July 2, 2025, the firm notified customers via Urgent Medical Device Recall letters. Customers were instructed to locate and quarantine any affected products at their facility and return an enclosed business reply form to the firm. Stryker will provide a shipping label to return the affected product, and credit will be issued to affected customers upon receipt of the recalled product.
- Quantity in commerce
- 233.230
- Distribution
- Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom
- Product codes and lots
- 5921-018-135 GTIN 07613154599257 5921-018-235 GTIN 07613154599295 5921-024-235 GTIN 07613154599370
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2277-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.