Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Boston Scientific Neuromodulation Corporation began a recall of Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09 on . FDA classified it as Class II and gives the reason as "Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2558-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Neuromodulation Corporation
- Product
- Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
- Product code
- NHL Stimulator, Electrical, Implanted, For Parkinsonian Symptoms
- Reason for recall
- Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On 7/8/2025, correction notices were mailed in the U.S. and mailed, emailed, or hand delivered outside the U.S. Healthcare Professionals were asked to do the following: 1) Review the Surgical Implant Manual updates. 2) If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. 3) A patient letter is enclosed with this communication, which can be shared with your patient and/or included within the patient's medical record. 4) Monitor per relevant IFU recommendations for any clinical observations of high monopolar impedances, undesired sensation, sudden loss of therapy, return of pre-implant symptoms and/or Bluetooth connectivity challenges, as these may be signals of potential feedthrough wire break(s) (component damage). 5) If you are a distributor, this notification must be forwarded to your customers to ensure this notification is carried out to the end-user level. 6) To provide awareness of this information, share this letter with any other clinicians in your hospital who use the system. 7) Maintain a copy of this letter in your facility's records. 8) Complete and return the acknowledgement form via email to [email protected] report all device-related incidents or quality concerns experienced with the use of these devices to the firm at [email protected]
- Quantity in commerce
- 25,260
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
- Product codes and lots
- UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2558-2025 | Class II | Ongoing | Voluntary: Firm initiated |