Z-2558-2025Class II

Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09

Boston Scientific Neuromodulation Corporation / initiated 2025-07-08 / Open, Classified / root cause: labeling and UDI

Boston Scientific Neuromodulation Corporation began a recall of Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09 on . FDA classified it as Class II and gives the reason as "Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2558-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Neuromodulation Corporation
Product
Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Product code
NHL Stimulator, Electrical, Implanted, For Parkinsonian Symptoms
Reason for recall
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 7/8/2025, correction notices were mailed in the U.S. and mailed, emailed, or hand delivered outside the U.S. Healthcare Professionals were asked to do the following: 1) Review the Surgical Implant Manual updates. 2) If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. 3) A patient letter is enclosed with this communication, which can be shared with your patient and/or included within the patient's medical record. 4) Monitor per relevant IFU recommendations for any clinical observations of high monopolar impedances, undesired sensation, sudden loss of therapy, return of pre-implant symptoms and/or Bluetooth connectivity challenges, as these may be signals of potential feedthrough wire break(s) (component damage). 5) If you are a distributor, this notification must be forwarded to your customers to ensure this notification is carried out to the end-user level. 6) To provide awareness of this information, share this letter with any other clinicians in your hospital who use the system. 7) Maintain a copy of this letter in your facility's records. 8) Complete and return the acknowledgement form via email to [email protected] report all device-related incidents or quality concerns experienced with the use of these devices to the firm at [email protected]
Quantity in commerce
25,260
Distribution
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
Product codes and lots
UDI-DI: 08714729985044, 08714729985051. Document Number/Revision: 92328632-09, Rev A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2558-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2558-2025Class IIOngoingVoluntary: Firm initiated