Z-2296-2025Class II

Custom Waste Management Kit: REF: K10-00353K, K10-01603C, K10-01694Q, K10-05159B, K10-05743, K10-05825AP

Merit Medical Systems, Inc. / initiated 2025-06-05 / Open, Classified / root cause: manufacturing process

Merit Medical Systems, Inc. began a recall of Custom Waste Management Kit: REF: K10-00353K, K10-01603C, K10-01694Q, K10-05159B, K10-05743, K10-05825AP on . FDA classified it as Class II and gives the reason as "High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2296-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Merit Medical Systems, Inc.
Product
Custom Waste Management Kit: REF: K10-00353K, K10-01603C, K10-01694Q, K10-05159B, K10-05743, K10-05825AP
Product code
PPD Universal Drainage Tray
Reason for recall
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 5/5/2025, recall notices were emailed to customers who were asked to do the following: 1) Identify affected devices and quarantine, discontinue use and distribution. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4. Return all affected devices to the firm. 5. Complete, and return the customer response form via email to [email protected] If you have questions, contact the firm's customer service at [email protected] or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri
Quantity in commerce
858
Distribution
US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.
Product codes and lots
REF/UDI-DI/Lot(Expiration): K10-00353K/00884450179246/H3127803(1/4/2028), H3166061(2/24/2028), H3172299(3/14/2028); K10-01603C/00884450593639/H3163984(3/2/2028), H3176777(3/23/2028); K10-01694Q/00884450239308/H3116890(10/31/2027), H3152493(2/3/2028), H3177622(3/14/2028); K10-05159B/00884450153598/H3116922(12/23/2027), H3119571(1/4/2028), H3127907(1/11/2028), H3152490(2/10/2028), H3152491(2/10/2028), H3152492(2/10/2028); K10-05743/00884450456767/H3122012(1/6/2028), H3128429(1/13/2028), H3177446(3/16/2028); K10-05825AP/00884450808108/H3112585(11/27/2027)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2296-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2293-2025Class IIOngoingVoluntary: Firm initiated
Z-2294-2025Class IIOngoingVoluntary: Firm initiated
Z-2295-2025Class IIOngoingVoluntary: Firm initiated
Z-2296-2025Class IIOngoingVoluntary: Firm initiated