Z-2194-2025Class II
LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)
Microbiologics, Inc. / initiated 2025-06-27 / Open, Classified
Microbiologics Inc began a recall of LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack) on . FDA classified it as Class II and gives the reason as "KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2194-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Microbiologics, Inc.
- Product
- LYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- KWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.
- FDA root cause
- Under Investigation by firm
- Action
- Microbiologics issued an Urgent Medical Device Recall notice on 06/27/2025 via email. The notice explained the problem with the device, potential risk, and requested the following: INSTRUCTIONS: 1. REVIEW lab procedures to understand how this information affects your usage. 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage. 3. COMPLETE the response form provided. 4. RETURN the response form to [email protected]. 5. KEEP this letter for your records. 6. NOTIFY other facilities or customers to whom the product may have been distributed or transferred. 7. CONTACT Microbiologics if a replacement kit is needed. Contact Recall Support team at 320.229.7080 or [email protected] with questions or concerns.
- Quantity in commerce
- 32 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of France, Colombia, Philippines, Peru, Turkey, India, Finland, Trinidad and Tobago, Poland, Viet Nam, Brazil, Japan, Malaysia, Costa Rica.
- Product codes and lots
- 1) 0894K: UDI/DI 30845357021954, Lot Numbers: 894-76-81; 2) 0894P: UDI/DI 20845357021940, Lot Numbers: 894-76-21, 894-76-51, 894-76-52.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2194-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2195-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.