Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
Capnia Inc began a recall of Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor on . FDA classified it as Class II and gives the reason as "Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2198-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Capnia Inc
- Product
- Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
- Product code
- CCJ Analyzer, Gas, Carbon-Monoxide, Gaseous-Phase
- Reason for recall
- Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986
- Quantity in commerce
- 2160
- Distribution
- US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.
- Product codes and lots
- Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2198-2025 | Class II | Ongoing | Voluntary: Firm initiated |