Z-2272-2025Class II

epoc BGEM BUN Test Card [25pk]. Material Number: 10736515

Siemens Healthcare Diagnostics Inc. / initiated 2025-07-07 / Open, Classified

Siemens Healthcare Diagnostics Inc began a recall of epoc BGEM BUN Test Card [25pk]. Material Number: 10736515 on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2272-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
FDA root cause
Under Investigation by firm
Action
Siemens Healthcare Diagnostics notified consignees via letter on about 07/07/2025. Consignees were instructed to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, discard and request a replacement of any affected lots or if your institution does not need to report sodium, you can turn off sodium following the instruction of the owner s manual. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form included with letter, notify customers if products were further distributed, and to retain this letter with your laboratory records and forward this letter to those who may have received this product at your site.
Quantity in commerce
37,554 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.
Product codes and lots
Material Number: 10736515. GTIN: 00809708121860. Lot Numbers: 02-25095-30, 02-25096-30, 02-25098-30, 02-25099-30, 02-25100-30, 02-25101-30, 02-25102-30, 02-25103-30, 02-25105-30, 02-25111-30, 02-25112-30, 02-25114-30, 02-25115-30, 02-25116-30, 02-25117-30, 02-25119-30, 02-25121-30, 02-25122-30, 02-25124-30, 02-25096-40, 02-25097-40, 02-25098-40, 02-25099-40, 02-25101-40, 02-25102-40, 02-25105-40, 02-25106-40, 02-25121-40, 02-25122-40, 02-25124-40, 02-25125-40, 02-25096-50, 02-25098-50, 02-25099-50, 02-25100-50, 02-25102-50, 02-25103-50, 02-25105-50, 02-25107-50, 02-25108-50, 02-25111-50, 02-25112-50, 02-25113-50, 02-25114-50, 02-25115-50, 02-25119-50, 02-25120-50, 12-25096-60, 12-25097-60, 12-25098-60, 12-25100-60, 12-25101-60, 12-25102-60, 12-25103-60, 12-25111-60, 02-25112-60, 02-25113-60, 02-25114-60, 02-25115-60, 02-25116-60, 02-25117-60, 02-25118-60, 02-25120-60, 02-25121-60, 02-25122-60, 02-25123-60, 02-25124-60, 02-25099-70, 02-25101-70, 02-25104-70, 02-25105-70, 02-25107-70, 02-25108-70, 02-25110-70, 02-25111-70, 02-25112-70, 02-25113-70, 02-25115-70, 02-25116-70, 02-25118-70, 02-25120-70.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2272-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2272-2025Class IIOngoingVoluntary: Firm initiated