Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2265-2025Class III

Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description...

DRG International, Inc. / initiated 2025-06-04 / Open, Classified / root cause: labeling and UDI

DRG International, Inc. began a recall of Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA on . FDA classified it as Class III and gives the reason as "An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., ...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2265-2025
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
DRG International, Inc.
Product
Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
Product code
CDX Radioimmunoassay, Total Thyroxine
Reason for recall
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).
Root cause tag
labeling and UDI
FDA root cause
Reprocessing Controls
Action
On June 4, 2025 FIELD SAFETY NOTICE letters were sent to customers. Action to be taken by distributors and end-users: Please check your inventory for the respective lot. Please use the CoA provided with this FSN. If you require corrected outer kit box label for your documentation, please contact DRG. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter. If you have any questions concerning this issue, please do not hesitate to contact us.
Quantity in commerce
47 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, PA and the countries of Germany, Hong Kong, Moldova.
Product codes and lots
Model/Catalog Number: EIA-4568; UDI-DI: 04048474045680; Lot number: 304K034;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2265-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2265-2025Class IIIOngoingVoluntary: Firm initiated