RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097
Siemens Healthcare Diagnostics Inc began a recall of RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097 on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2230-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- Siemens Healthcare sent an Urgent Medical Device Correction (UMDC) letter to consignees (via FedEx in the US) on around 06/26/2025. The communication letter informs consignees of the issue and provided actions for consignees to take. Consignees were instructed to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, replace the both the measurement cartridge and the wash/waste cartridge if still using a wash/waste cartridge from lot WW/08925 with RAPIDPoint systems, discontinue use of the wash/waste cartridges from lot WW/08925 and dispose any remaining inventory in possession. No-cost replacement wash/waste cartridges should be made through the local Siemens Healthineers or distributor office. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form included with letter, notify customers if products were further distributed, and to retain this letter with your laboratory records and forward this letter to those who may have received this product at your site.
- Quantity in commerce
- 1,477 units
- Distribution
- worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
- Product codes and lots
- Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.