Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2167-2025Class I

Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with...

Hamilton Medical AG / initiated 2025-06-26 / Open, Classified / root cause: manufacturing process

Hamilton Medical AG began a recall of Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators on . FDA classified it as Class I and gives the reason as "Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2167-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Hamilton Medical AG
Product
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Product code
BZO Set, Tubing And Support, Ventilator (W Harness)
Reason for recall
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Starting on 6/26/2025, recall notices were emailed to customers who were asked to do the following: Return affected devices. If one of the affected breathing circuits would be already in use with a patient, it is requested to exchange it and discard the product after replacement. If you are a distributor, please forward to any organization where potentially affected product has been transferred. Complete and return the acknowledgement form via email to [email protected] If you have any questions contact firm at: [email protected] and 800-426-6331
Quantity in commerce
2,560
Distribution
US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.
Product codes and lots
UDI-DI: 07630002802963, Lots: 199675, 199676

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2167-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2167-2025Class IOngoingVoluntary: Firm initiated