Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0272-2026Class II

Model Number SM-40HF-B-D-C; 40KW 70C EN, Mobile X-ray system

Sedecal S.A. / initiated 2025-06-10 / Open, Classified / root cause: design

SEDECAL SA began a recall of Model Number SM-40HF-B-D-C; 40KW 70C EN, Mobile X-ray system on . FDA classified it as Class II and gives the reason as "Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0272-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Sedecal S.A.
Product
Model Number SM-40HF-B-D-C; 40KW 70C EN, Mobile X-ray system
Product code
IZL System, X-Ray, Mobile
Reason for recall
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
Root cause tag
design
FDA root cause
Device Design
Action
Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.
Quantity in commerce
1 unit
Distribution
US: CA, IL, and NJ
Product codes and lots
UDI/DI 08436046001510, Serial Numbers: G62066

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0272-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0242-2026Class IIOngoingVoluntary: Firm initiated
Z-0243-2026Class IIOngoingVoluntary: Firm initiated
Z-0244-2026Class IIOngoingVoluntary: Firm initiated
Z-0245-2026Class IIOngoingVoluntary: Firm initiated
Z-0246-2026Class IIOngoingVoluntary: Firm initiated
Z-0247-2026Class IIOngoingVoluntary: Firm initiated
Z-0248-2026Class IIOngoingVoluntary: Firm initiated
Z-0249-2026Class IIOngoingVoluntary: Firm initiated
Z-0250-2026Class IIOngoingVoluntary: Firm initiated
Z-0251-2026Class IIOngoingVoluntary: Firm initiated
Z-0252-2026Class IIOngoingVoluntary: Firm initiated
Z-0253-2026Class IIOngoingVoluntary: Firm initiated
Z-0254-2026Class IIOngoingVoluntary: Firm initiated
Z-0255-2026Class IIOngoingVoluntary: Firm initiated
Z-0256-2026Class IIOngoingVoluntary: Firm initiated
Z-0257-2026Class IIOngoingVoluntary: Firm initiated
Z-0258-2026Class IIOngoingVoluntary: Firm initiated
Z-0259-2026Class IIOngoingVoluntary: Firm initiated
Z-0260-2026Class IIOngoingVoluntary: Firm initiated
Z-0261-2026Class IIOngoingVoluntary: Firm initiated
Z-0262-2026Class IIOngoingVoluntary: Firm initiated
Z-0263-2026Class IIOngoingVoluntary: Firm initiated
Z-0264-2026Class IIOngoingVoluntary: Firm initiated
Z-0265-2026Class IIOngoingVoluntary: Firm initiated
Z-0266-2026Class IIOngoingVoluntary: Firm initiated
Z-0267-2026Class IIOngoingVoluntary: Firm initiated
Z-0268-2026Class IIOngoingVoluntary: Firm initiated
Z-0269-2026Class IIOngoingVoluntary: Firm initiated
Z-0270-2026Class IIOngoingVoluntary: Firm initiated
Z-0271-2026Class IIOngoingVoluntary: Firm initiated
Z-0272-2026Class IIOngoingVoluntary: Firm initiated
Z-0273-2026Class IIOngoingVoluntary: Firm initiated
Z-0274-2026Class IIOngoingVoluntary: Firm initiated
Z-0275-2026Class IIOngoingVoluntary: Firm initiated
Z-0276-2026Class IIOngoingVoluntary: Firm initiated
Z-0277-2026Class IIOngoingVoluntary: Firm initiated
Z-0278-2026Class IIOngoingVoluntary: Firm initiated
Z-0279-2026Class IIOngoingVoluntary: Firm initiated
Z-0280-2026Class IIOngoingVoluntary: Firm initiated
Z-0281-2026Class IIOngoingVoluntary: Firm initiated
Z-0282-2026Class IIOngoingVoluntary: Firm initiated
Z-0283-2026Class IIOngoingVoluntary: Firm initiated
Z-0284-2026Class IIOngoingVoluntary: Firm initiated
Z-0285-2026Class IIOngoingVoluntary: Firm initiated
Z-0286-2026Class IIOngoingVoluntary: Firm initiated
Z-0287-2026Class IIOngoingVoluntary: Firm initiated