Model Number SM-40HF-B-D-C; 40KW 70C READY, Mobile X-ray system
SEDECAL SA began a recall of Model Number SM-40HF-B-D-C; 40KW 70C READY, Mobile X-ray system on . FDA classified it as Class II and gives the reason as "Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0273-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Sedecal S.A.
- Product
- Model Number SM-40HF-B-D-C; 40KW 70C READY, Mobile X-ray system
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.
- Quantity in commerce
- 10 units
- Distribution
- US: CA, IL, and NJ
- Product codes and lots
- UDI/DI 08436046001510, Serial Numbers: G37158, G61698, G62863, G41192, G62814, G41098, G62566, G60400, G63392, G60406.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0242-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0243-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0244-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0245-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0246-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0247-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0248-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0249-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0250-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0251-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0252-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0253-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0254-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0255-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0256-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0257-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0258-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0259-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0260-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0261-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0262-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0263-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0264-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0265-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0266-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0267-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0268-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0269-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0270-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0271-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0272-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0273-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0274-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0275-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0276-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0277-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0278-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0279-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0280-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0281-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0282-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0283-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0284-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0285-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0286-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0287-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.