Z-2338-2025Class II

Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy...

OLYMPUS Corporation of the Americas / initiated 2025-07-25 / Open, Classified

Olympus Corporation of the Americas began a recall of Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410... on . FDA classified it as Class II and gives the reason as "Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2338-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
FDA root cause
Component change control
Action
Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.
Quantity in commerce
155 units
Distribution
US Nationwide distribution.
Product codes and lots
Model Number: WA91327U. Lot Code: ESG-410 (SEP) WA91327U. UDI-DI: 04042761087698. Serial Numbers: 101696, 101697, 101698, 101699, 101700, 101701, 101702, 101703, 101704, 101705, 101706, 101707, 101708, 101709, 101710, 101919, 101920, 101921, 101922, 101923, 101924, 101925, 101926, 101927, 101928, 101929, 101933, 101934, 101935, 101936, 101937, 101940, 101941, 101942, 101945, 101946, 101947, 102148, 102149, 102150, 102151, 102152, 102153, 102154, 102156, 102157, 102158, 102159, 102160, 102161, 102162, 102164, 102165, 102167, 102168, 102170, 102172, 102173, 102174, 102175, 102176, 102177, 102178, 102179, 102180, 102181, 102182, 102183, 102184, 102185, 102186, 102187, 102188, 102189, 102190, 102191, 102193, 102194, 102195, 102198, 102199, 102200, 102266, 102427, 101932, 102155, 102163, 102197, 102660, 102661, 102662, 102663, 102664, 102665, 102666, 102667, 102668, 102669, 102670, 102671, 102672, 102673, 102674, 102675, 102676, 102677, 102678, 102679, 102680, 102681, 102682, 102683, 102684, 102685, 102686, 102687, 102688, 102689, 102690, 102691, 102692, 102693, 102694, 102695, 102696, 102697, 102698, 102699, 102700, 102701, 102702, 102703, 102704, 102705, 102706, 102707, 102708, 102709, 102710, 102711, 102712, 102713, 102714, 102715, 102716, 102717, 102718, 102719, 102721, 102722, 102723, 102724, 102725, 103922, 103930.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2338-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2338-2025Class IIOngoingVoluntary: Firm initiated
Z-2339-2025Class IIOngoingVoluntary: Firm initiated