Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number...
Olympus Corporation of the Americas began a recall of Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is... on . FDA classified it as Class II and gives the reason as "Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2339-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
- FDA root cause
- Component change control
- Action
- Olympus notified consignees via letter on about 07/25/2025. Consignees were instructed to examine inventory, identify any affected units, and continued use is permitted until the repair is performed as long as users follow the IFU. Olympus will contact consignees to arrange for the return of affected units to replace the affected part. Consignees were also instructed to notify customers, if the units were further distributed, and to acknowledge receipt of the letter through the Olympus web portal.
- Quantity in commerce
- 434 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model Number: WA91307C. Lot Code: ESG-410 Legacy WA91307C. UDI-DI: 04042761086349. Serial Numbers: 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101023, 101024, 101025, 101029, 101030, 101031, 101032, 101033, 101034, 101035, 101036, 101037, 101038, 101039, 101040, 101041, 101042, 101043, 101044, 101045, 101046, 101047, 101048, 101049, 101050, 101051, 101052, 101053, 101054, 101055, 101056, 101057, 101058, 101059, 101060, 101061, 101062, 101063, 101064, 101066, 101067, 101068, 101069, 101070, 101071, 101072, 101073, 101074, 101238, 101239, 101240, 101241, 101242, 101243, 101244, 101245, 101246, 101247, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101260, 101261, 101262, 101263, 101264, 101265, 101266, 101267, 101268, 101269, 101270, 101271, 101272, 101273, 101274, 101275, 101276, 101277, 101278, 101279, 101280, 101281, 101282, 101283, 101284, 101285, 101286, 101287, 101288, 101289, 101290, 101291, 101292, 101293, 101294, 101295, 101296, 101297, 101298, 101299, 101300, 101301, 101302, 101333, 101334, 101335, 101336, 101337, 101338, 101339, 101340, 101341, 101342, 101370, 101371, 101372, 101373, 101374, 101375, 101376, 101377, 101378, 101379, 101395, 101396, 101397, 101398, 101399, 101400, 101401, 101402, 101403, 101432, 101433, 101434, 101435, 101436, 101437, 101438, 101439, 101440, 101441, 101613, 101633, 101634, 101635, 101636, 101637, 101638, 101684, 101685, 101686, 101687, 101688, 101689, 101691, 101690, 101692, 101693, 101694, 101695, 101743, 101744, 101745, 101746, 101747, 101748, 101749, 101750, 101751, 101752, 101753, 101754, 101755, 101756, 101757, 101758, 101759, 101760, 101761, 101762, 101763, 101764, 101765, 101766, 101767, 101768, 101769, 101770, 101771, 101772, 101773, 101774, 101775, 101776, 101777, 101778, 101779, 101780, 101781, 101782, 101783, 101784, 101785, 101786, 101787, 101788, 101789, 101958, 101959, 101960, 101961, 101962, 101963, 101964, 101965, 101966, 101967, 101968, 101969, 101970, 101971, 102017, 102018, 102019, 102020, 102021, 102024, 102025, 102026, 102027, 102028, 102055, 102056, 102057, 102058, 102059, 102060, 102061, 102062, 102063, 102064, 102065, 102066, 102067, 102068, 102069, 102284, 102285, 102286, 102287, 102288, 102289, 102290, 102291, 102292, 102293, 102294, 102295, 102296, 102297, 102298, 102299, 102300, 102301, 102302, 102303, 102304, 102305, 102306, 102307, 102308, 102309, 102310, 102311, 102312, 102313, 102314, 102315, 102316, 102317, 102318, 102319, 102320, 102321, 102323, 102325, 102326, 102327, 102328, 102329, 102330, 102331, 102332, 102333, 102334, 102335, 102362, 102363, 102364, 102366, 102367, 102368, 102369, 102370, 102371, 102372, 102373, 102391, 102392, 102393, 102394, 102395, 102396, 102397, 102399, 102400, 102401, 102402, 102403, 102404, 102405, 102406, 102407, 102408, 102409, 102410, 102411, 102412, 102413, 102414, 102415, 102416, 102417, 102418, 102419, 102420, 102421, 102422, 102423, 102424, 102425, 102426, 102428, 102429, 102430, 102431, 102432, 102433, 102434, 102435, 102436, 102437, 102438, 102439, 102440, 102441, 102442, 102443, 102444, 102445, 102497, 102498, 102499, 102500, 102501, 102502, 102503, 102504, 102505, 102506, 102507, 102508, 102509, 102510, 102511, 102512, 102513, 102515, 102516, 102517, 102518, 102519, 102520, 102521, 102522, 102523, 102802, 102807, 102808, 102809, 102810, 102814, 102818, 102819, 102820, 102821, 102822, 102823, 102828, 102829, 102830, 102831, 102832, 102834, 102836, 102837.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2338-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2339-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY