InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)
Medtronic MiniMed, Inc. began a recall of InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users) on . FDA classified it as Class II and gives the reason as "Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2496-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)
- Product code
- NDC Calculator, Drug Dose
- Reason for recall
- Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- On 6/16/2025, correction notices were emailed to customers who were asked to do the following: Actions Required: Users of InPen app software should update to version 7.5.1 or higher. For iOS Users - Verify your app version: - On the home screen, navigate to the Settings menu (lower right corner). - Select Help and Support. - Note the App Version displayed. - Open the App Store. - Tap your profile icon at the top of the screen. - Scroll to see pending updates. - Tap Update next to the InPen app icon. For Android Users - Verify your app version: - On the home screen, navigate to the Settings menu (lower right corner). - Select Help and Support. - Note the App Version displayed. - Open the Google Play Store. - Tap your profile icon at the top of the screen, then select Manage apps & device. - Tap Updates available or search for the InPen app. - Tap Update next to the InPen app icon. Firm's 24-Hr Technical Support: 1-800-646-4633
- Quantity in commerce
- 6816
- Distribution
- International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
- Product codes and lots
- UDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2496-2025 | Class II | Ongoing | Voluntary: Firm initiated |