Injection needle, bayonet, Luer-Lock, 1. Model no. 50-353-23, working length 25 cm, total length 31cm...
Spiggle & Theis Mt Gmbh began a recall of Injection needle, bayonet, Luer-Lock, 1. Model no. 50-353-23, working length 25 cm, total length 31cm, nonsterile; 2. Model no. 50-345-23, working length 24... on . FDA classified it as Class II and gives the reason as "Affected product show elevated cytotoxicity values. Use of affected injection needles may present a risk of intolerance reaction or allergic reaction". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2331-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Spiggle & Theis Medizintechnik GmbH
- Product
- Injection needle, bayonet, Luer-Lock, 1. Model no. 50-353-23, working length 25 cm, total length 31cm, nonsterile; 2. Model no. 50-345-23, working length 24 cm, total length 31 cm, non-sterile
- Product code
- GDM Needle, Aspiration And Injection, Reusable
- Reason for recall
- Affected product show elevated cytotoxicity values. Use of affected injection needles may present a risk of intolerance reaction or allergic reaction.
- FDA root cause
- Under Investigation by firm
- Action
- On June 27, 2025, the firm notified customers through a letter titled "Urgent safety information Recall Injection Needle". Customers were asked to identify and quarantine any affected product in stock. Customer service will assist customers in returning product and the firm will issue a credit upon receipt of the affected product.
- Distribution
- US Nationwide distribution in the states of California, Connecticut, Missouri, and Ohio.
- Product codes and lots
- 1. Model No 50-353-23; UDI-DI 04250381858806; LOT (0)2241239 2. Model No 50-345-23; UDI-DI 04250381858813; LOT 2230438; LOT 2240251
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2331-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY