SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A
C.R. Bard Inc began a recall of SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A on . FDA classified it as Class II and gives the reason as "Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2589-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- C.R. Bard Inc
- Product
- SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A
- Product code
- MJC Catheter, Urological (Antimicrobial) And Accessories
- Reason for recall
- Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- On 7/17/2025, recall notices were mailed and emailed to Recall Coordinator, Director of Nursing, Director of Purchasing, Director of Risk Management, and Distributors who were asked to do the following: 1) It is recommended that clinicians discontinue the use of impacted devices and find alternative catheters with the appropriate characteristics required for their patients to continue therapy and avoid potential health risks related to this issue. 2) Quarantine and discard all devices within your facility s control per your facility s procedures. 3) Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Recall (Correction). 4) If you have purchased this product through a distributor, please contact this distributor to request appropriate credit or order alternate replacement product. 5) Complete and return the attached Customer Response Form via email to [email protected] In addition, distributors were asked to: a) Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification on the recalling firm's behalf. b) Contact recalling firm if you require assistance with credits to be issued for discard-ed products or for replacement product options. If you require assistance contact the firm: Medical Information Services: 1-800-555-7422, [email protected]; Technical Support: 1-844-823-5433, [email protected]
- Quantity in commerce
- 4,300
- Distribution
- US Nationwide distribution including in the states of NC, FL, NY, MA, MN, TX, CA, OH, MI, WA, NJ, IL, ID, SC, IN, OK, OR, AL, VA, HI, MO, MS, MD, AZ, KY, GA, IA, CO, NM, PA, ME, UT, NE.
- Product codes and lots
- Lot#NGJX6399
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2589-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2590-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2591-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.