Z-2620-2025Class II

Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034

STRYKER Communications, Inc. / initiated 2025-07-15 / Open, Classified

Stryker Communications began a recall of Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 on . FDA classified it as Class II and gives the reason as "Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2620-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
STRYKER Communications, Inc.
Product
Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
FDA root cause
Component change control
Action
On 7/15/2025, recall notices were mailed or hand delivered to customers who were asked to do the following: Inform individuals within your organization who need to be aware of this action. 2. A Stryker representative will be in contact with your account to schedule a field service visit at your earliest convenience to complete the following: a. Deinstall impacted Oculan light(s) and reinstall with SLX lights to ensure your operating room stays in use. b. Upon product remediation and availability, Stryker will return to reinstall Oculan within your operating room(s). Your local Stryker representative will provide timing updates as they are available. 3. Complete and return the Business Reply Form to [email protected] 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Quantity in commerce
40
Distribution
US Nationwide distribution in the states of TX, IL, NY, UT, OR, AZ, VT, NY, IL, MN.
Product codes and lots
UDI-DI: 07613327296167 and 07613327651317 REF/Serial Numbers: CH00000001/7000010116, 7000010117, 7000010118, 7000010119, 7000010120, 7000010121, 7000010122, 7000010123, 7000010124, 7000010125, 7000010141, 7000010142, 7000010177, 7000010178, 7000010148, 7000010149, 7000010157, 7000010137, 7000010175, 7000010176, 7000010145, 7000010146, 7000010147, 7000010309, 7000010305, 7000010307, 7000010296, 7000010143, 7000010150, 7000010151, 7000010155. P60034/1000022, 1000023, 1000024, 1000016, 1000017, 1000018, 1000019, 1000020, 1000021

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2620-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2620-2025Class IIOngoingVoluntary: Firm initiated