FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro...

BioFire Defense, LLC / initiated 2025-07-10 / Open, Classified / root cause: design

Biofire Defense began a recall of FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0... on . FDA classified it as Class II and gives the reason as "FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2609-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
BioFire Defense, LLC
Product
FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system.
Product code
PRD
Reason for recall
FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.
Root cause tag
design
FDA root cause
Component design/selection
Action
On 7/10/2025, correction notices were mailed and emailed to customers who were asked to do the following: - If testing with only whole blood and/or sputum, there is no impact to the product s performance. Continue using panel kits as planned. - If testing with positive blood culture may occur, discontinue use, dispose of impacted kit(s) according to your lab s procedures, and only use replacement kits provided, or use impacted lots only with whole blood and/or sputum sample types. - Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties who may use this product, including others to whom you may have transferred our product. - Complete and return the acknowledgment of receipt form via email to [email protected]
Quantity in commerce
130 Kits
Distribution
US Nationwide distribution in the states of AL, AZ, UT, CA, TX, DC , VA, FL, MD, OH, HI, NY, KY, LA, MA, MI, MN, MO, NC, NH, RI, NV, TN, IA, WA.
Product codes and lots
UDI-DI: 00851458005136. Lot(Expiration): 324324D(2026-02-19), D241022(2026-01-22), 319524D(2025-11-20), 315424D(2025-10-09), 315224D(2025-10-03), 315324D(2025-10-08)