Z-2540-2025Class I

Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCMC use in conjunction with the...

Permobil, AB / initiated 2025-08-12 / Open, Classified / root cause: design

Max Mobility LLC began a recall of Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCMC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component on . FDA classified it as Class I and gives the reason as "Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2540-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Permobil, AB
Product
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCMC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
Product code
ITI Wheelchair, Powered
Reason for recall
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
Root cause tag
design
FDA root cause
Component design/selection
Action
On August 12, 2025, Permobil issued a "Urgent Medical Device Recall Expansion/Update" via email. This expansion supersedes the Class I recall ((Z-1116-2025, Z-1117-2025), initiated on December 20, 2024. Permobil asked consignees to take the following actions: 1. Immediately discontinue use of the SpeedControl dial to reduce the likelihood of a potentially hazardous or harmful situation. 1. Since all SpeedControl Dials are impacted by this recall, remove SpeedControl Dials from your sellable / distribution inventory 2. Use an alternate method to control the SmartDrive MX2+ Power Assist Device. Alternate controls include: " A wired SwitchControl, which can be ordered with a mono jack port to connect with an optional Buddy Button " Wirelessly through the SmartDrive MX2+ App utilizing a Apple iWatch or Samsung Galaxy watch. 3.Immediately contact any customers you have distributed the device to and provide a copy of this notice. It is also recommended that this notice be posted in your clinic/facility 4. You are required to acknowledge that you received this Urgent Medical Device Recall letter. Please complete within 10 days of receipt of the letter. Please use the QR code below to access the field action portal to record your acknowledgement
Quantity in commerce
8,413 units
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: CAN, NZL, KOR, BMU, GBR, JPN, TWN, AUS, BRA, COL, MEX, ABW, SWE, CHL, ARG, HKG
Product codes and lots
All Serial Numbers distributed between the dates of April 25, 2022 - July 8, 2025

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2540-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2538-2025Class IOngoingVoluntary: Firm initiated
Z-2539-2025Class IOngoingVoluntary: Firm initiated
Z-2540-2025Class IOngoingVoluntary: Firm initiated