TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number...

Premia Spine Ltd. / initiated 2025-07-16 / Open, Classified

PREMIA SPINE LTD began a recall of TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889 on . FDA classified it as Class II and gives the reason as "Potential for missing pins at tip of inserter". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2516-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Premia Spine Ltd.
Product
TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.
Product code
QWK Prosthesis, Posterior Spinal Elements
Reason for recall
Potential for missing pins at tip of inserter.
FDA root cause
Process change control
Action
On 07/16/2025, Premia Spine Ltd. notified Premia Spine Inc. via phone call. It was requested to quarantine any suspected reworked instruments that are at the warehouse and to initiate a field action and return the affected units of the TOPS Inserter that underwent rework. No customers (hospitals/physicians) were asked to quarantine or return units as all affected units were within Premia Spine's control, however one surgeon was notified, on about 07/31/2025, of the issue and was requested to review the surgical procedure, analyze relevant patient data, including x-ray, and perform additional actions if needed to confirm the pin has not remained in the patient's body.
Quantity in commerce
30 units
Distribution
US Nationwide distribution in the state of Connecticut.
Product codes and lots
Model Number: 82889. UDI-DI: 07290115932638. Lot Numbers: RS20220546-1, RS20222162-1, RS20171777-1, RS20171884-1, RS0431-13B-1, 20152764-1, RS0431-02-1.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2516-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2516-2025Class IIOngoingVoluntary: Firm initiated