Z-2573-2025Class II
UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
Siemens Healthcare Diagnostics Inc. / initiated 2025-08-13 / Open, Classified
Beckman Coulter Inc. began a recall of UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520 on . FDA classified it as Class II and gives the reason as "Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2573-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- UniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- FDA root cause
- Component change control
- Action
- On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System (LIS), Middleware, or on affected instruments. Consult your Medical Director to determine if a retrospective review of results is warranted. Beckman Coulter recommends posting this letter on or near the affected analyzers Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. Acknowledge receipt of the notification.
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.
- Product codes and lots
- UDI: 15099590729110/ Serial Numbers: BD17734, BE12048, BE28105, BF14045, BF23096, BF46236, BG26135, BG50247, BH12019, BH43168
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2572-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2573-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2574-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2575-2025 | Class II | Ongoing | Voluntary: Firm initiated |