Z-0036-2026Class II

GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A...

W. L. Gore & Associates, Inc. / initiated 2025-08-26 / Open, Classified / root cause: labeling and UDI

W L Gore & Associates, Inc. began a recall of GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A... on . FDA classified it as Class II and gives the reason as "Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0036-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
W. L. Gore & Associates, Inc.
Product
GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use.
Product code
DSY Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Reason for recall
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
Root cause tag
labeling and UDI
FDA root cause
Unknown/Undetermined by firm
Action
On 8/26/2025, correction notices were mailed to Vascular Surgeons and other physicians who were informed of the following: No action is required for patients who currently have the device implanted; however, please be aware of delamination as a potential cause of occlusion should clinical issues arise in the future. Management of patients with suspected or confirmed delamination should be guided by the clinical judgment of the treating physician, with careful consideration of the risks and benefits of each treatment option in the context of the individual circumstance. Instructions for use will be updated to include the following: When performing reinterventions on the GORE ACUSEAL Vascular Graft, avoid the use of oversized balloons. A possible complication which may occur with the use of the GORE ACUSEAL Vascular Graft is delamination. Delamination may result in partial or complete occlusion due to hemodynamically significant stenosis or thrombosis and related serious harms, including additional interventions to resolve. IFUs are available at: https://eifu.goremedical.com/ In addition, customers were asked to do the following: 1) Complete and return the acknowledgement form via email to [email protected] 2) This notice needs to be shared with those who should be aware within your institution or to any organization where potentially affected devices have been transferred (as appropriate). To report an event to firm, email: [email protected] Questions can be directed to firm Customer Service (email: [email protected] or by phone at 800-528-8763
Quantity in commerce
52,306
Distribution
US: OR, AL, OH, MD, VA, CA, FL, NC, OK, MA, MO, AZ, TX, IL, GA, PA, NH, HI, MN, MS, KY, ID, LA, NJ, TN, DE, IN, MI, WI, NY, DC, SC, CT, NE, IA, WA, AR, NV, WV, AK, UT, ND, KS, CO, ME, MT, NM, SD, PR, WY, RI. OUS: PANAMA, CANADA, CHINA, TAIWAN, GERMANY, SLOVENIA, NORWAY, UNITED KINGDOM, ITALY, SPAIN, FINLAND, NETHERLANDS, SWEDEN, FRANCE, UNITED ARAB EMIRATES, DENMARK, PORTUGAL, AUSTRIA, SWITZERLAND, BELGIUM, TURKIYE, SLOVAK REPUBLIC, CYPRUS, SAUDI ARABIA, SOUTH AFRICA, LUXEMBOURG, FRENCH GUIANA, ANDORRA, IRELAND, NEW CALEDONIA, POLAND, ISRAEL, GREECE, GUADELOUPE, CROATIA, REUNION, BAHRAIN, HUNGARY, JAPAN
Product codes and lots
REF/UDI-DI: ECH050020J/04993024010758, ECH050020W/00733132635450, ECH050050J/04993024010772, ECH050050W/00733132635474, ECH060010A/00733132622573, ECH060020A/00733132622597, ECH060020J/04993024009967, ECH060020W/00733132634729, ECH060040/00733132605347, ECH060040A/00733132605354, ECH060040W/00733132619719, ECH060050A/00733132626960, ECH060050J/04993024009981, ECH460045A/00733132634811, ECH460045J/04993024010789, ECH470045/00733132626977, ECH470045A/00733132634811. Product with a serial number that includes and falls within the range of 6597081PP001 and 9853822PP024. Instructions for Use MD183888 Revision 4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0036-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0036-2026Class IIOngoingVoluntary: Firm initiated