Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and...
Medline Industries, LP began a recall of Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as... on . FDA classified it as Class I and gives the reason as "Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2614-2025
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. For imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. Medline Item Numbers 10135910RH 10135936RH
- Product code
- OWQ Reprocessed Intravascular Ultrasound Catheter
- Reason for recall
- Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
- FDA root cause
- Under Investigation by firm
- Action
- On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.
- Quantity in commerce
- 7913 eaches
- Distribution
- Domestic US distribution nationwide. International distribution pending. No international distribution.
- Product codes and lots
- Item Number/UDI-DI 10135910RH (for use on GE Systems) 10197344026500; 10135936RH (For use on Siemens Systems) 10197344026517; LOTS EP241021 EP241114 EP241226 EP250213 LOTS EP240731 EP240814 EP240826 EP240903 EP240916 EP241003 EP241014 EP241028 EP241104 EP241112 EP241202 EP241209A EP241219 EP241224 EP250114 EP250116 EP250204 EP250219
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2610-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2611-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2612-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2613-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2614-2025 | Class I | Ongoing | Voluntary: Firm initiated |