Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645...
LeMaitre Vascular, Inc. began a recall of Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740... on . FDA classified it as Class II and gives the reason as "The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed and approved by the appropriate regulatory authority". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0034-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- LeMaitre Vascular Inc.
- Product
- Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030;
- Product code
- LXA Tissue Graft Of 6mm And Greater
- Reason for recall
- The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed and approved by the appropriate regulatory authority.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On September 2, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Check your inventory against the list of lots in this letter. Immediately quarantine any recalled devices. 2. Complete and return attached response form. 3. If the product has been implanted, graft explantation/patient intervention is NOT required. 4. When a recalled device has been returned to LeMaitre Vascular, a replacement device will be provided. 5. If you have transferred devices to another facility, please forward a copy of this recall letter to them. If you have any questions, contact Monte Nelson; [email protected]; 732-422-8333 Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.
- Quantity in commerce
- 28 units
- Distribution
- US Nationwide distribution in the states of CA, CT, FL, GA, IL, MA, MS, NC, ND, NJ, NY, OK, PA, TN, TX, VA, WA.
- Product codes and lots
- Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030; UDI-DIs: (1) 00316837000213, (2) 00316837000220, (3) 00316837000237, (4) 00316837000244, (5) 00316837000251, (6) 00316837000268, (7) 00316837000275, (8) 00316837000282, (9) 00316837000299, (10) 00316837000305, (11) 00316837000312, (12) 00316837000367, (13) 00316837000329, (14) 00316837000343, (15) 00316837000350; Serial Numbers: (1) 23DD203-019, 23DD203-046, (2) 23DD203-024, 23DD203-029, (3) 23DD203-045, (4) 23DD203-018, (5) 23DD203-004, (6) 23DD203-021, 23DD203-044, 23DD203-038, 23DD203-039, 23DD203-011, 23DD203-032, 23DD203-031, 23DD203-043, (7) 23DD203-040, 23DD203-030, 23DD203-037, (8) 23DD203-035, 23DD203-033, (9) 23DD203-034, (10) 23DD203-036, (11) 23DD203-022, 23DD203-023, (12) 23DD203-017, (13) 23DD203-027, (14) 23KK498-014, (15) 23DD203-026;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0034-2026 | Class II | Ongoing | Voluntary: Firm initiated |