Z-2618-2025Class II

Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm...

Boston Scientific Corporation / initiated 2025-08-28 / Open, Classified / root cause: manufacturing process

Boston Scientific Corporation began a recall of Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above) on . FDA classified it as Class II and gives the reason as "The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2618-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below), (2) M00547000 (9-12mm Above);
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On August 28, 2025 URGENT MEDICAL DEVICE REMOVAL letters were sent to customers. Instructions: 1. Immediately cease further distribution or use of any remaining affected devices. Segregate the units in a secure place until they can be returned. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. 4. Return affected devices.
Quantity in commerce
156 units
Distribution
US Nationwide distribution in the states of CA, CO, GA, ID, IL, IN, MA, ME, MI, MS, NC, OH, OK, PA, SD, TN, TX, VA, WA.
Product codes and lots
Instructions for Use: 51878205; Pre-Printed Ouch: 51368417; 1. UPN (REF): M00547030 (9-12mm Below); UDI-DI: 08714729790280; Lot Numbers: 36398057; 2. UPN (REF): (1) M00547000(9-12mm Above); UDI-DI: 08714729790266; Lot Numbers: 36426495;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2618-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2618-2025Class IIOngoingVoluntary: Firm initiated