Z-0372-2026Class II

SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Purple, Model: 07 6011...

ASP Global, LLC / initiated 2025-05-05 / Open, Classified / root cause: design

ASP Global, LLC. dba Anatomy Supply Partners, LLC. began a recall of SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 L... on . FDA classified it as Class II and gives the reason as "All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0372-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
ASP Global, LLC
Product
SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL End-to-End Capillary Tubes, 30 L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Root cause tag
design
FDA root cause
Component design/selection
Action
Starting on 5//5/2025 customers were informed that there had been reports of falsely elevated blood lead levels associated with the use of affected micro capillary tubes for lead testing when used with Magellan LeadCare Testing Systems. On 9/8/2025, recall notices were emailed to customers who were asked to do the following: Discontinue use of the recalled products immediately. b) Quarantine any remaining inventory to prevent further use. c) Dispose remaining inventory in accordance with local regulations or notify firm at [email protected] for return authorization and disposal assistance. d) Complete and return the enclosed Recall Response Form to confirm receipt and actions to [email protected] Distributors are asked to notify their customers of the recall or to forward their customer information to the recalling firm so they might notify these customers.
Quantity in commerce
4,272,858
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
Product codes and lots
Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0372-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0372-2026Class IIOngoingVoluntary: Firm initiated