Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2670-2025Class I

Medline Kits: 1) VALVE PACK-LF, Model Number: DYNJ0415366P; 2) VALVE PACK-LF, Model Number: DYNJ0415366Q

MEDLINE INDUSTRIES, LP - Northfield / initiated 2025-08-21 / Open, Classified

Medline Industries, LP began a recall of Medline Kits: 1) VALVE PACK-LF, Model Number: DYNJ0415366P; 2) VALVE PACK-LF, Model Number: DYNJ0415366Q on . FDA classified it as Class I and gives the reason as "The kits contain certain lots of cannula products where the catheter may not retain its shape". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2670-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Kits: 1) VALVE PACK-LF, Model Number: DYNJ0415366P; 2) VALVE PACK-LF, Model Number: DYNJ0415366Q
Product code
OEZ Cardiovascular Procedure Kit
Reason for recall
The kits contain certain lots of cannula products where the catheter may not retain its shape.
FDA root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE RECALL notification dated 8/21/25 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com. Recall Reference #: R-25-181-FGX1 Recall Code: 3. Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Quantity in commerce
828 units
Distribution
US Nationwide distribution in the state of CA.
Product codes and lots
1) DYNJ0415366P, UDI-DI: 10195327177966(each), 40195327177967(case), Lot Number: 24DMG366; 2) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 25EMD672; 3) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 25GMD016; 4) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 25AMB610; 5) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 25BMJ241; 6) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 25CMI425; 7) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 24DMC189; 8) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 24EMC463; 9) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 24GMJ353; 10) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 24IMB824; 11) DYNJ0415366Q, UDI-DI: 10195327678227(each), 40195327678228(case), Lot Number: 24KME116

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2670-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2670-2025Class IOngoingVoluntary: Firm initiated