MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI...

Elekta Solutions AB / initiated 2025-09-17 / Open, Classified / root cause: software

ELEKTA SOLUTIONS AB began a recall of MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1... on . FDA classified it as Class II and gives the reason as "When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0503-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Elekta Solutions AB
Product
MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0 MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.
Root cause tag
software
FDA root cause
Software design
Action
On 09/17/2025, the firm emailed an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" to customers informing them when appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule. Customers are instructed to: When appending cycles for care plans that contain additional medication orders with a wave schedule, review and adjust the appended care plan cycles to reflect the clinical intent. Alternatively, wave orders can be removed from the care plan folder. Additional information can be found in: Customers with v3.1.3 - section 3.4.1 - page 35-36 in doc E062133 / 01 MOSAIQ 3.1.3.0 Instructions for Use Customers with v3.2.1 - section 3.4.1 - page 35-36 in doc E065117 / 03 MOSAIQ 3.2.1.0 Instructions for Use Customers with v3.2.2 - section 3.4.1 - page 35-36 in doc E067652 / 04 MOSAIQ 3.2.2.0 Instructions for Use Customers with v3.2.3.0 - section 3.4.1 - page 35-36 in doc E071967 / 01 MOSAIQ 3.2.3.0 Instructions for Use For questions - contact Quality Assurance at +46 70-962-0862 or email [email protected]
Quantity in commerce
4 systems that contain the affected software.
Distribution
Worldwide - U.S. Nationwide distribution in the state of NM and the countries of Canada, Malta, and South Africa.
Product codes and lots
Software Versions: 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 UDI codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500026(10)3.1.3.2 3.1.3.3 (01)07340201500026(10)3.1.3.3 3.1.3.4 (01)07340201500026(10)3.1.3.4 3.2.1.0 (01)07340201500071(10)3.2.1.0 3.2.1.1 (01)07340201500071(10)3.2.1.1 3.2.1.2 (01)07340201500071(10)3.2.1.2 3.2.1.3 (01)07340201500071(10)3.2.1.3 3.2.1.4 (01)07340201500071(10)3.2.1.4 3.2.2.0 (01)07340201500071(10)3.2.2.0 3.2.2.1 (01)07340201500071(10)3.2.2.1 3.2.3.0 (01)07340201500071(10)3.2.3.0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0503-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0503-2026Class IIOngoingVoluntary: Firm initiated