FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
FUJIFILM Healthcare Americas Corporation began a recall of FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33 on . FDA classified it as Class II and gives the reason as "It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0044-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On September, 15, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Perform the daily inspection described in the operation manual to check for wobbles or shakes, misalignment of the light irradiation position, etc. If any abnormality is found, discontinue use and contact Technical Assistance Center (TAC). When rotating the X-ray tube unit, do not hit the stopper at the end. Do not rotate the X-ray tube over or near the patient. Please post this notice on or near the device. Complete and return the response form provided. Fujifilm Service Engineers will install countermeasure parts that will prevent the X-ray tube support from coming off even if the shaft is damaged. Please contact Technical Assistance Center (TAC) at 1-888-385-4633 or 1-888-FUJIMED for any functional/correction questions as well as to schedule your correction.
- Quantity in commerce
- 16 units
- Distribution
- Domestic: AK, CA, FL, IL, MA, NH, OH, WI.
- Product codes and lots
- Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0044-2026 | Class II | Ongoing | Voluntary: Firm initiated |