Z-0071-2026Class II

SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only

Siemens Healthcare Diagnostics Inc. / initiated 2025-09-03 / Open, Classified / root cause: manufacturing process

Beckman Coulter Inc. began a recall of SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only on . FDA classified it as Class II and gives the reason as "Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0071-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
Product code
CEO Phosphomolybdate (Colorimetric), Inorganic Phosphorus
Reason for recall
Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
Root cause tag
manufacturing process
FDA root cause
Unknown/Undetermined by firm
Action
On September 3, 2025, Beckman Coulter issued a "Urgent Medical Device Recall" Notification to affected consignees via E-Mail. Beckman Coulter asked consignees to take the following actions: 1. You may continue to use the affected lots if calibration passes and quality control results are within your facility s acceptance criteria. 2. If the above lots fail calibration due to INIT ADC HI before the lot expiration date, contact Customer Technical Support Center for troubleshooting and/or replacement. 3. If you receive replacement product, review the Safety Data Sheet (SDS) and discard the affected lot(s). 4. Beckman Coulter offers SYNCHRON Systems Cartridge Chemistry Phosphorus (PHS REF A09426) Reagent as an alternative. 5. Please share this information with your laboratory staff. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 6. Please respond within 10 days.
Quantity in commerce
2146 units
Distribution
US: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico Singapore
Product codes and lots
UDI: 15099590233358/ Lot Numbers: M309642 M309641 M401490 M404365 M406380 M406417 M408408 M410466 M410513 M412278

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0071-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0071-2026Class IIOngoingVoluntary: Firm initiated