Salem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux...
Cardinal Health 200, LLC began a recall of Salem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 266130 Salem Sump"... on . FDA classified it as Class II and gives the reason as "Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0210-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Salem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 266130 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) 266148 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 18 Fr/Ch 8888266106 Salem Sump" Dual Lumen Stomach Tube, Anti-Reflux Valve 10Fr/Ch (3.3 mm), 36" 266130CN Salem Sump" Tube 16FR with CONFIRM NOW, Nonsterile 266148CN Salem Sump" Tube 8 FR with CONFIRM NOW, Nonsterile
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- FDA root cause
- Use error
- Action
- On September 18, 2025, the firm notified customers via Urgent Medical Device Product Correction Update letters. This is an update to the notice sent in July 2024 for RES 95214. The present recall includes all lots of additional product codes listed. The letter provides customers with additional guidance on how to properly use the Anti-Reflux Valve (ARV): Important Instructions: After placement of the Salem Sump tube, always inject 10ml-20ml of air before seating the Anti Reflux Valve (ARV) into the blue sump vent lumen. Instructions: 1. Seat the BLUE END of the ARV into the blue sump vent lumen of the Salem Sump tube. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 2. To cap tube, seat the WHITE END of the ARV into the suction lumen. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 3. To remove, pull valve in the same direction of assembly. Do not pull at an angle to avoid breaking the valve. Please visit https://www.mycardinalmsds.com for updated electronic IFUs.
- Quantity in commerce
- all lots
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
- Product codes and lots
- Product Code/Product Name/UDI-DI: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 10884521065024 266130 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) 10884521065031 266148 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 18 Fr/Ch 10884521065055 8888266106 Salem Sump" Dual Lumen Stomach Tube, Anti-Reflux Valve 10Fr/Ch (3.3 mm), 36" 20884521004655 266130CN Salem Sump" Tube 16FR with CONFIRM NOW, Nonsterile 10884521141292 266148CN Salem Sump" Tube 8 FR with CONFIRM NOW, Nonsterile 20884521141305 All Lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0206-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0207-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0208-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0209-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0210-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0211-2026 | Class II | Ongoing | Voluntary: Firm initiated |