Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0318-2026Class I

Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM...

Dexcom, Inc. / initiated 2025-08-28 / Open, Classified / root cause: software

Dexcom, Inc. began a recall of Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software... on . FDA classified it as Class I and gives the reason as "A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0318-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Dexcom, Inc.
Product
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System
Product code
QBJ Integrated Continuous Glucose Monitoring System, Factory Calibrated
Reason for recall
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).
Root cause tag
software
FDA root cause
Software Design Change
Action
On 09/18/2025, the firm sent via in-app and email an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" communication informing customer that a bug has been identified within the G6 Android app version 1.15.0 that can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, or alerts. Customers are instructed to: Upgrading from app version 1.15.0 is mandatory and they will not be able to use app version 1.15.0 after October 6th, 2025. Update your app now to continue using the app: 1. Open the App Store and search for the Dexcom G6 app. 2. Install the latest Dexcom G6 app version. 3. Open the Dexcom G6 app. For questions or further assistance - contact Technical Support at 1-844-478-1600. Technical Support is available 24 hours a day/7 days a week
Quantity in commerce
333
Distribution
This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.
Product codes and lots
Lot Code: UDI: SW11678 - 00386270000811

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0318-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0318-2026Class IOngoingVoluntary: Firm initiated