Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning...
Kico Knee Innovation Company began a recall of Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product... on . FDA classified it as Class II and gives the reason as "Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0145-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Kico Knee Innovation Company
- Product
- Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.
- Product code
- SBF Orthopedic Augmented Reality
- Reason for recall
- Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On September 12, 2025, Kico Knee Innovation Company issued a recall notice to affected consignees. Kico asked consignees to take the following actions: 1. No actions are required by the surgeon users. 2. For the plans affected, Kico has quarantined these as they are not suitable for clinical use. 3. A long-term solution has been implemented to return the system to the specification of the most recently cleared device. We will contact surgeon users to login and re-plan their cases for upcoming surgeries.
- Quantity in commerce
- 21 systems
- Distribution
- US Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY, CA, WI, MI, PA, WA, CO, AL.
- Product codes and lots
- Software Version V2025.1.2