Z-0134-2026Class II

Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen...

Iantrek Inc. / initiated 2025-09-06 / Open, Classified / root cause: manufacturing process

Iantrek Inc. began a recall of Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual... on . FDA classified it as Class II and gives the reason as "Potential for compromised integrity of the sterile packaging". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0134-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Iantrek Inc.
Product
Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual surgical instrument used for the ab-interno construction or modification of a cyclodialysis, followed by delivery of viscous materials, bio-tissue/allograft, or other materials to reinforce and maintain the cleft.
Product code
HMX Cannula, Ophthalmic
Reason for recall
Potential for compromised integrity of the sterile packaging.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Iantrek notified consignees on 09/06/2025 via email. Consignees were instructed to immediately remove and quarantine all affected units to prevent their use, notify personnel or customers if affected units were forwarded to other facilities. Consignees may also alternatively provide Iantrek with contact information so that they can notify them on the customer's behalf. An Iantrek representative will be contacting each consignee immediately to help facilitate removal, return and replacement of affected units. They will arrange to have replacement product shipped to consignees at no cost.
Quantity in commerce
1,262 units
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.
Product codes and lots
1. Model Number CP-300; UDI: +B962CP3000/$$7351645/16D20250201/14D20260201/Q1_; Lot Number: 351645; Exp date 02/2026. 2. Model Number CP-300; UDI: +B962CP3000/$$7351726/16D20250501/14D20260501/Q11; Lot Number 351726; Exp date 05/2026. 3. Model Number CP-300; UDI: +B962CP3000/$$7351795/16D20250601/14D20260601/Q19; Lot Number 351795; Exp date 06/2026.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0134-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0134-2026Class IIOngoingVoluntary: Firm initiated
Z-0135-2026Class IIOngoingVoluntary: Firm initiated