Z-0288-2026Class II
HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J
HeartSine Technologies Ltd / initiated 2025-09-18 / Open, Classified
HeartSine Technologies Ltd began a recall of HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J on . FDA classified it as Class II and gives the reason as "Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0288-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- HeartSine Technologies Ltd
- Product
- HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
- FDA root cause
- Under Investigation by firm
- Action
- On September 18, 2025, HeartSine (Stryker) issued a "Urgent: Medical Device Recall Notification." HeartSine is asking consignees to take the following actions: 1. To ensure the device works correctly in an emergency, follow the instructions within the customer notification. A. Check the expiry date on your Pad-Pak B. Check for bent locator pins. C. Follow Pad-Paks insertion instructions. D. Submit your response online at: heartsincerecall.com. A response is required in all instances. E. Maintain Awareness F. If you have further distributed affected product(s), please forward a copy of this notice to the new responsible party and email [email protected] with both the location to which the product(s) was/were further distributed to and the quantity and lot number for any Pad-Paks that have been disposed of. If you have any questions or concerns, please contact Stryker Customer Service at +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday, or HeartSine Technologies Technical Support at [email protected].
- Quantity in commerce
- 1,291,165 units
- Distribution
- US Nationwide distribution. Worldwide distribution to Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Product codes and lots
- PAD-PAK-01, Lot: H017-002-174 and H047-002-100, UDI: 0506016712046; PAD-PAK-02, Lot: H023-002-051, UDI: 0506016712047; PAD-PAK-03, Lot: H017-002-173, H047-002-101, UDI: 0506016712123; PAD-PAK-04, Lot: H023-002-050, UDI: 0506016712124; PAD-PAK-07, Lot: H017-002-178, UDI: 0506016712448; PAD-PAK-03J, Lot: H017-002-071 and H047-002-035, UDI: 0506016712475; PAD-PAK-04J, Lot: H023-002-009, UDI: 0506016712476
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0288-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY