Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps...
Tandem Diabetes Care, Inc. began a recall of Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin... on . FDA classified it as Class I and gives the reason as "Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0427-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Tandem Diabetes Care, Inc.
- Product
- Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081
- Product code
- QFG Alternate Controller Enabled Insulin Infusion Pump
- Reason for recall
- Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 10/6/2025, correction notices were mailed or emailed to customers who were informed the following: If your pump shows Malfunction 12, insulin delivery will stop, and the pump becomes non-operational: 1) Promptly acknowledge it by pressing "Silence Alarm". 2) Contact Tandem Diabetes Care Customer Technical Support (CTS) via phone at 1-877-801-6901 or email at [email protected] for further assistance. 3) Switch to your backup method of insulin delivery as directed by your physician. 4) Regularly check your blood sugar to ensure you are not having unexpectedly high or low readings. Update your pump software as soon as possible to version 7.9.0.2. You can check your software version within the Mobi Mobile App. Select Settings > Pump > Pump Info. The latest software can be downloaded remotely from within the Tandem Mobi Mobile App. Once logged in to the app, tap "Settings" and then "Pump" to view software updates. Complete the acknowledgement form using link: https://campaign.tandemdiabetes.com/FCA-Malfunction-12 User Guide: https://www.tandemdiabetes.com/docs/default-source/user-guide/user-guide-tandem-mobi-control-iq-7-9-mgdl-en-us-aw1014938.pdf?sfvrsn=872fc4d7_10
- Quantity in commerce
- 17,745
- Distribution
- US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY
- Product codes and lots
- Software versions: 7.6.0.1, 7.6.0.3, and 7.7.0.1 Catalog/UDI-DI: 1010750/00389152075013, 1012719/00389152271910, 1013501/00389152350110, 1013655/00389152365510, 1013656/00389152365619, 1013700/00389152370019, 1014081/00389152408170
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0427-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.