Z-0452-2026Class II

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number...

OLYMPUS Corporation of the Americas / initiated 2025-10-08 / Open, Classified / root cause: design

Olympus Corporation of the Americas began a recall of Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse... on . FDA classified it as Class II and gives the reason as "The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0452-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System
Product code
FEO Lithotriptor, Ultrasonic
Reason for recall
The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.
Root cause tag
design
FDA root cause
Device Design
Action
On October 8, 2025 URGENT: MEDICAL DEVICE ADVISORY NOTICE letters were sent to customers. Actions to be taken: Ensure you read the content of the notice. Ensure that all personnel are aware and knowledgeable on the content of the notice. You may continue to use the device according to the instruction for use, which cautions users to ensure that a back-up transducer and probe are sterilized and available prior to beginning a procedure. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you acknowledge receipt of this letter with the provided reply form.
Quantity in commerce
1528 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.
Product codes and lots
Model/Catalog Number: SPL-T; UDI-PI: 00855279005023; All serial numbers;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0452-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0452-2026Class IIOngoingVoluntary: Firm initiated