Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0170-2026Class II

Basic Biopsy Tray, Item Number NMBP44L

American Contract Systems, Inc. / initiated 2025-09-04 / Open, Classified / root cause: manufacturing process

American Contract Systems Inc. began a recall of Basic Biopsy Tray, Item Number NMBP44L on . FDA classified it as Class II and gives the reason as "Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0170-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
American Contract Systems, Inc.
Product
Basic Biopsy Tray, Item Number NMBP44L
Product code
FCH Tray, Biopsy, Without Biopsy Instruments
Reason for recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Quantity in commerce
32,433 total
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Product codes and lots
UDI-DI 191072188120 Lot 8515411 Entire lot impacted

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0170-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0170-2026Class IIOngoingVoluntary: Firm initiated
Z-0171-2026Class IIOngoingVoluntary: Firm initiated
Z-0172-2026Class IIOngoingVoluntary: Firm initiated
Z-0173-2026Class IIOngoingVoluntary: Firm initiated
Z-0174-2026Class IIOngoingVoluntary: Firm initiated
Z-0175-2026Class IIOngoingVoluntary: Firm initiated
Z-0176-2026Class IIOngoingVoluntary: Firm initiated
Z-0177-2026Class IIOngoingVoluntary: Firm initiated
Z-0178-2026Class IIOngoingVoluntary: Firm initiated
Z-0179-2026Class IIOngoingVoluntary: Firm initiated
Z-0180-2026Class IIOngoingVoluntary: Firm initiated
Z-0181-2026Class IIOngoingVoluntary: Firm initiated
Z-0182-2026Class IIOngoingVoluntary: Firm initiated
Z-0183-2026Class IIOngoingVoluntary: Firm initiated
Z-0184-2026Class IIOngoingVoluntary: Firm initiated
Z-0185-2026Class IIOngoingVoluntary: Firm initiated
Z-0186-2026Class IIOngoingVoluntary: Firm initiated
Z-0187-2026Class IIOngoingVoluntary: Firm initiated