Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0176-2026Class II

Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR...

American Contract Systems, Inc. / initiated 2025-09-04 / Open, Classified / root cause: manufacturing process

American Contract Systems Inc. began a recall of Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK... on . FDA classified it as Class II and gives the reason as "Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0176-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
American Contract Systems, Inc.
Product
Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA75I VASCULAR PACK
Product code
OEZ Cardiovascular Procedure Kit
Reason for recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
Quantity in commerce
32,433 total
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Product codes and lots
ANCV78BB UDI-DI 191072214362 Lot 8133811 Bag serial number 68273380; ANCV78BC UDI-DI 191072235169 Lot 8499811 Bag serial numbers N/A; HGCV05K UDI-DI 191072212474 Lot 8021411 Bag serial number 68154464; MHPP97AR UDI-DI 191072237446 Lot 8527511 Bag serial number N/A; UDBH37AW UDI-DI 191072235299 Lot 8194311 Bag serial numbers 68198334 68198335; UHPP82AH UDI-DI 191072216434 Lot 8112811 Bag serial number N/A; UICD62AN UDI-DI 191072200495 Lot 7999011 Bag serial number N/A; UICT16F UDI-DI 191072150431 Lot 8000411 Bag serial number N/A Lot 8331311 Bag serial number 17627819 Lot 8429011 Bag serial number 17749736; UIVA75I UDI-DI 191072229052 Lot 8056011 Bag serial number N/A; UIVA75J UDI-DI 191072235602 Lot 8422011 Bag serial number 68667536 * If Bag serial number is "n/a", then all of that lot is affected

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0176-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0170-2026Class IIOngoingVoluntary: Firm initiated
Z-0171-2026Class IIOngoingVoluntary: Firm initiated
Z-0172-2026Class IIOngoingVoluntary: Firm initiated
Z-0173-2026Class IIOngoingVoluntary: Firm initiated
Z-0174-2026Class IIOngoingVoluntary: Firm initiated
Z-0175-2026Class IIOngoingVoluntary: Firm initiated
Z-0176-2026Class IIOngoingVoluntary: Firm initiated
Z-0177-2026Class IIOngoingVoluntary: Firm initiated
Z-0178-2026Class IIOngoingVoluntary: Firm initiated
Z-0179-2026Class IIOngoingVoluntary: Firm initiated
Z-0180-2026Class IIOngoingVoluntary: Firm initiated
Z-0181-2026Class IIOngoingVoluntary: Firm initiated
Z-0182-2026Class IIOngoingVoluntary: Firm initiated
Z-0183-2026Class IIOngoingVoluntary: Firm initiated
Z-0184-2026Class IIOngoingVoluntary: Firm initiated
Z-0185-2026Class IIOngoingVoluntary: Firm initiated