Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK...
American Contract Systems Inc. began a recall of Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH... on . FDA classified it as Class II and gives the reason as "Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0180-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224
- Product code
- OEQ Angiography/Angioplasty Kit
- Reason for recall
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
- Quantity in commerce
- 32,433 total
- Distribution
- US Nationwide distribution in the states of SD, IA, MN, WA, IL.
- Product codes and lots
- ANCA80AP UDI-DI 191072232168 Lot 8077111 Bag serial number 68119740; HSAN63N UDI-DI 191072232267 Lot 8097211 Bag serial number N/A; IHAN02W UDI-DI 191072201928 Lot 7961811 Bag serial number 68086881 Lot 8037411 Bag serial number N/A; MHCA98AK UDI-DI 191072216014 Lot 8006711 Bag serial number 47415134 Lot 8133011 Bag serial number N/A; RCAN45C UDI-DI 191072229632 Lot 8537111 Bag serial number 68872542; UDCO10F UDI-DI 191072216236 Lot 7987811 Bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0170-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0171-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0172-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0173-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0174-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0175-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0176-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0177-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0178-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0179-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0180-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0181-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0182-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0183-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0184-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0185-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0186-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0187-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.