Z-0420-2026Class II

TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number)...

Integra LifeSciences Corporation / initiated 2025-10-03 / Open, Classified / root cause: design

Integra LifeSciences Corp. (NeuroSciences) began a recall of TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01 on . FDA classified it as Class II and gives the reason as "It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0420-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Integra LifeSciences Corporation
Product
TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
Product code
PGW Ear, Nose, And Throat Stereotaxic Instrument
Reason for recall
It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
Root cause tag
design
FDA root cause
Device Design
Action
On October 3, 2025 URGENT: VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken by Customers 1. Examine the your system to determine if you have product on hand subject to this field recall (Rev 00 and Rev01). 2. If you have revision 00 and/or revision 01, please check the box on the enclosed Acknowledgement Form I do have affected product(s) and record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the rest of the Acknowledgement Form and return. 5. Keep a copy of the form for your records. 6. Forward this notice to those who utilize the product so they are aware of this recall and can identify and quarantine any affected product that may remain in clinical areas. 7. When your form is received, and it is noted that you have Rev 00 or Rev 01 products, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. You can request a replacement which you will receive depending on availability. 8. In an effort to prioritize replacement Adapters please note below (in the Medical Facilities Acknowledgement Form) if you have Rev 02 and Rev 03 to use for your upcoming planned surgeries.
Quantity in commerce
1198 units
Distribution
US Nationwide distribution.
Product codes and lots
Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0420-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0420-2026Class IIOngoingVoluntary: Firm initiated