Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0316-2026Class II

Magseed Pro 7 cm soft tissue marker. Model Number: MS2-17-1-07. Product Description: The Endomag Magseed Pro...

Endomagnetics Ltd / initiated 2025-10-01 / Open, Classified / root cause: manufacturing process

ENDOMAGNETICS LTD began a recall of Magseed Pro 7 cm soft tissue marker. Model Number: MS2-17-1-07. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use... on . FDA classified it as Class II and gives the reason as "Potential for contamination with cotton fibers". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0316-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Endomagnetics Ltd
Product
Magseed Pro 7 cm soft tissue marker. Model Number: MS2-17-1-07. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
Product code
NEU Marker, Radiographic, Implantable
Reason for recall
Potential for contamination with cotton fibers.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Endomagnetics, Ltd notified consignees on about 10/01/2025 via certified mail and email through 3rd party IQVIA. Consignees were instructed to examine their inventory and cease use of those units, quarantine any affected units, complete and return the response form, forward the notification to all affected personnel and customers if further distributed, and post a copy of the notice in a visible area for awareness and keep a copy for your records.
Quantity in commerce
1,760 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, CO, CT, DC, FL, GA, IA, ID, IN, KY, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, TX, WA and the country of Canada.
Product codes and lots
Model Number. MS2-17-1-07. GTIN:15060391210459. UDI: 15060391210459241227-01;15060391210459250212-02; 15060391210459250324-03; 15060391210459250415-06; 15060391210459250523-11; 15060391210459250612-51 Batch Numbers: 250212-02, 250324-03, 250415-06, 250523-11, 241227-01, 250612-51.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0316-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0316-2026Class IIOngoingVoluntary: Firm initiated
Z-0317-2026Class IIOngoingVoluntary: Firm initiated