Z-0579-2026Class II

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V...

OLYMPUS Corporation of the Americas / initiated 2025-10-14 / Open, Classified / root cause: design

Olympus Corporation of the Americas began a recall of Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended... on . FDA classified it as Class II and gives the reason as "Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0579-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
Product code
FDT Duodenoscope And Accessories, Flexible/Rigid
Reason for recall
Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
Root cause tag
design
FDA root cause
Device Design
Action
On October 14, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Carefulle read the notification and ensure that all personnel are completely knowledgeable and thoroughly trained on the content of the letter. Complete the required training provided in the letter. Keep a copy of the notification with the operation manual. It is recommended to receive on-site TJF duodenoscope reprocessing training at your facility. If you have further distributed this product, identify and forward them this notification.
Quantity in commerce
7023 units
Distribution
US Nationwide distribution.
Product codes and lots
Model/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0579-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0579-2026Class IIOngoingVoluntary: Firm initiated