Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V...
Olympus Corporation of the Americas began a recall of Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended... on . FDA classified it as Class II and gives the reason as "Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0579-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.
- Product code
- FDT Duodenoscope And Accessories, Flexible/Rigid
- Reason for recall
- Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On October 14, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Carefulle read the notification and ensure that all personnel are completely knowledgeable and thoroughly trained on the content of the letter. Complete the required training provided in the letter. Keep a copy of the notification with the operation manual. It is recommended to receive on-site TJF duodenoscope reprocessing training at your facility. If you have further distributed this product, identify and forward them this notification.
- Quantity in commerce
- 7023 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0579-2026 | Class II | Ongoing | Voluntary: Firm initiated |