Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0...
Spacelabs Healthcare, Ltd. began a recall of Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management... on . FDA classified it as Class II and gives the reason as "Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0416-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 09/15/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them of two issues associated with the Sentinel Software Version 11.6.0. The issues are: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record. Customers are instructed to: 1. Print and Post Notification: Please print the related Urgent User Notification (Appendix A) in all location(s) where Sentinel 11.6.0 is used to configure ABP and/or Ambulatory ECG (Holter) devices. 2. Notify all relevant employees within their organization, including satellites, of this recall 3. Confirm the above action has been taken: Go to the QR code below (Appendix B) and confirm that they have read, understood, and posted the user notification at their workstations. 4. Contact Customer Support team at [email protected] with any questions regarding this identified issue. A fix for these issues is being developed, and is anticipated to be available by the end of September 2025. A Spacelabs Healthcare representative will notify customers once the fix is available and will work with customers to schedule the software update. Note, rolling back Sentinel 11.6.0 to a previous version is not a valid option.
- Quantity in commerce
- 142
- Distribution
- Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.
- Product codes and lots
- Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0416-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.