Low Prof Lckng Screw, F/im Nail 5.0Mm / L, 46Mm / Xl25/ Ster
Synthes (USA) Products LLC began a recall of LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and... on . FDA classified it as Class II and gives the reason as "Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0228-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Synthes (USA) Products LLC
- Product
- LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
- Product code
- JDS Nail, Fixation, Bone
- Reason for recall
- Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- DePuy Synthes notified consignees via email and mailed letter on 09/26/2025. Consignees were instructed to examine inventory and quarantine affected units, arrange for the return of any affected units, complete and return the Business Response Form, and place the notification in a visible area for awareness. If units were further distributed within your facility or to customers, please notify those customers and personnel.
- Quantity in commerce
- 120 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
- Product codes and lots
- Product Code: 04.045.346TS. GTIN: 10886982301413. Lot Number: 31756P7. Expiration Date: 08/31/2025
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0228-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0229-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.