Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0229-2026Class II

Low Prof Lckng Screw, F/im Nail 5.0Mm / L, 42Mm / Xl25/ Ster

Synthes (USA) Products LLC / initiated 2025-09-26 / Open, Classified / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and... on . FDA classified it as Class II and gives the reason as "Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0229-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Synthes (USA) Products LLC
Product
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Product code
JDS Nail, Fixation, Bone
Reason for recall
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
DePuy Synthes notified consignees via email and mailed letter on 09/26/2025. Consignees were instructed to examine inventory and quarantine affected units, arrange for the return of any affected units, complete and return the Business Response Form, and place the notification in a visible area for awareness. If units were further distributed within your facility or to customers, please notify those customers and personnel.
Quantity in commerce
122 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
Product codes and lots
Product Code: 04.045.342TS. GTIN: 10886982301390. Lot Number: 31756P9. Expiration Date: 08/31/2025

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0229-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0228-2026Class IIOngoingVoluntary: Firm initiated
Z-0229-2026Class IIOngoingVoluntary: Firm initiated