Z-0154-2026Class I

IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23"...

Abbott Medical / initiated 2025-10-02 / Open, Classified / root cause: manufacturing process

ICU Medical, Inc. began a recall of IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off)... on . FDA classified it as Class I and gives the reason as "IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0154-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014
Product code
FMG Stopcock, I.V. Set
Reason for recall
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
ICU notified consignees via email with an "URGENT MEDICAL DEVICE CORRECTION" on about 10/02/2025. Consignees were instructed to: 1) Check all inventory locations within your institution for the affected products and discontinue use. Destroy all affected products following your institution s process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2) Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there. 3) Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. 4) If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. ICU Medical will provide replacement products to affected customers upon receipt of a complete Customer Response Form to certify product destruction. If you have any questions, contact Customer Service via email: [email protected] or call 1-(800)-258-5361 Additional information or technical assistance.
Quantity in commerce
23203
Distribution
Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
Product codes and lots
REF/UDI-DI/Lot: B33359/10887709007823/14070544, 14126963; B9213/10840619045859/14130197; B9732/10887709015033/14149636; B9733/10887709015040/14176779; 011-C7014/10840619005068/14029638, 14065820, 14110075, 14211300, 14129823

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0154-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0154-2026Class IOngoingVoluntary: Firm initiated